These equipments are used for climatic & durability tests of electrical & electronic components, corrosion test on mechanical assemblies, tests on packages, paints & varnishes, cement plants, apparatus & materials for simulated tropical & extreme tropical conditions & for botanical & zoological investigations, testing of chemicals, drugs, pharmaceutical, cosmetics, medicine, capsules etc.
Chamber runs on the basis of Dewpoint control. Less deviations during short-term door opening & power failure. Less utility area Increased storage area. Low operating cost. Minimum power & water consumption. Reduced deviations. Less breakdowns due to standby refrigeration & humidification system. Due to PLC HMI & controlling through microprocessor based PID Temp. & RH operations the on time of both Temp. & RH system is reduced & efficiency of this system is increased with reduced energy consumption.
Single Phase 230V AC, 50Hz.
MOC of the Chamber will be: Double walled Chamber with Inner of
stainless steel 304 grade & outer of CRCA duly powder coated,
PUF Insulation in between two walls.
(I) Double door, inner door of glass with frame & silicon gasket sealing,
outer door metallic with PUF insulation.
(ii) The unit will be made of double walled leak proof metal door with lock,
with a provision of opening door from inside.
(iii) Forced air circulation for uniform temperature.
(iv) Lockable & Movable PU wheels for easy movement.
Insulation of Walls, ceiling and floor : 80mm thick puff panels with
puff in-place density of minimum of 1kg per cubic foot. Exterior metal
surface of insulation puff panels or its equivalent shall be of minimum 19
gauge embossed white stainless steel or its equivalent. Interior metal
surface of insulation puff panels or its equivalent shall be of minimum 19
gauge stainless steel with baked white enamel finish or its equivalent.
Exterior and interior metal surface shall not have metal-to- metal bonding.
Door : Flush type doors of metal surfaces (as per above specification of
metal) with proper puff insulation of above mentioned standard or its
equivalent. Doors shall have magnetic snap-in perimeter gasket or its
equivalent, self-closing cam lift gravity hinges or its equivalent, a posi-seal
door closure or its equivalent. Door shall have key lockable latch handle.
Doorjambs shall be made fiberglass reinforced plastic or any other suitable
material compatible with constructed door
The chamber is provided with best air circulation to give better temperature uniformity this is achieved by using one set of motor & blower in baffle wall Compartment to push treated air in working chamber. This system gives better than 1 deg c. temperature uniformity in working chamber, evaporator, heaters & steam injection system is from back side, behind baffle wall, the air circulation blower is also fitted in this compartment only. This arrangement helps to mix steam and air together and pass well mixed air inside the working chamber. Due to above system the temperature and humidity (R.H.) in complete chamber almost remains equal.
First time in India developed by SR Lab Instruments.
Standby Refrigeration, Standby Humidification & Standby safety
controller with auto changeover, In case of failure of the main
system, To run the system for a longer with uninterrupted
operation.
Also USFDA approved 21 CFR compliance software, where in
case of temp. & RH fluctuation, SMS & Email alert shall be sent
to registered user nos and minute details of operation of the
chamber can be downloaded weekly, Monthly or Quarterly.
Standard PID:Provides Precise Temperature, Maintain Uniform
And Accurate Growth Environment, Timer, Alarm, Auto-Tuning
And Auto Start Stop Function. Technical Design and Construction
Silent Fan Motor to Maintain Uniform Temperature Over Temp.
and Current Protection Capability Ensure User Safety, Adjustable
Sliding Rack, Alert Unexpected Interruption Of Electrical Power
Or Unauthorized Change.
[Optional : ]
PLC with HMI Control system with touch screen display & HMI
for data Storage, data logger reading & audit trail, Door access
& SMS records, Email alert. Data Logger 1 temp + 1 RH. Data
Logger with LCD display window capable to store non volatile
data memory up to 5000 readings with audit trail facility complies
21 CFR Part 11. Change over to stand by system automatically
as per schedule or on detecting fault with current system Tempe.
& Humidity overshoot and under shoot protection Mobile Alarm
system, Door Access, system setting reset , Graph, SMS, Email
alert, etc. Data logger with RS 485/232 port for logging & transfer
of data. Auto restart, power failure event logging, memory
protection S.M.A.R.T/ SELF diagnostics system. Alert
Unexpected Interruption of Electrical Power or Unauthorized
Change.4 digit password lock.
Emerson / Danfoss / Tecumseh: make compressors, 1 stage cooling system for TEMP. & Humidity Control. All accessories such as HPCO/LPCO, Oil Separators, Dryers, Relays, OLP etc. of Danfoss/Tecumseh Emerson Make. CFC free refrigerant to be used to comply with the New Environmental Regulation. Will have suitable tonnage compressor for maintain temperature with lights on. Will have hermetically sealed condensing unit compressor coupled with Evaporation coil and condenser, safe guarded by time delay circuit. Cabinet shall be supplied with air-cooled condensing unit with extended compressor life and close temperature control. Cooling system will be split type to place condenser unit in a ventilated area to avoid hot air in growth chamber. Cooled air shall be delivered in such a way to reduce or eliminate condensation within media dishes and to provide maximum uniformity In different research vessels. This may be through silent and durable fan motors or through any kind of advanced air diffusers. Air Cooled Condenser with solenoid & LP to pass hot gas to the chamber for uniformity & long life of compressor for continuous sage. Phenolic coating for cooling & condensing coils (All weather proof).
+10°C to +60°C
Resolution : 0.1°C, Accuracy : 0.2°C, Uniformity : ± 2°C
Humidity Range : 40% to 98% RH,
Resolution : 0.1%, Accuracy : ± 1%, Uniformity : ± 3%.
Custom built stainless steel tubular heaters are used as heating element. The stainless steel fins ensure better heat transfer. Humidity is controlled by steam injection method. The boiler tank is used for steam generation; Electro-magnetic switch is used for sensing the Water Level. Thus eliminating the humidity interruption. De-humidification coils are installed below the cooling coils for lower humidity.
Sensors : Radix / Rotronic Make Hygroclip non condensing type sensor for RH and PT-100 ‘A’ class for temp. calibrated from ERTL / NABL accredited labs.
2 minute compressor “on” delay timer to safeguard the
Compressors overload relay protector. Electronic low water level
cut off device to cut off the supply to boiler heater in case of low
water level. Safety Temp. Controller. MCB for mains.
(I) Unit will be provided with safety devices for temperature and
humidity overshoot in case of malfunction.
(ii) Built-in temp. Deviation, audio/visual alarms.
Safety thermostat for over shooting of temp. Safety circuit to
cut off the whole system.
(iii) OLP (Overload Protector) & Time delay Circuit for safety of
compressors.
(iv) HRC fuses for compressors, Heaters & Mains.
ISO 9001:2015 Certified Co. D&B registered & CE mark product.
ISO 14001:2015 & WHO-GMP
NSIC & MSME registered
Calibration & validation certificates traceable to NABL / ERTL
accredited labs.
To comply with the documentation requirements, we provide
IQ, OQ & PQ protocols to be executed before taking the chamber
in to for regular use, and support the supply with the following
documents with the detail operational and service manual.
Standard Operating Procedure (SOP)
Operational Manual for Controller
IQ,PQ,OQ protocols certificate.
Calibration certificate of all controlling modules with traceability.
Certificate of MOC.
Test Report of chamber prior to supply with mapping certificate.
wiring diagram for ease of service maintenance.