Walk In Hot And Cold Chamber : Model No : SRL / WIHCCH 02
Walk In Hot And Cold Chamber :
Model No : SRL-WIHCCH 02
Application :
These equipments are used for climatic & durability tests of electrical & electronic components, corrosion test on mechanical assemblies, materials for simulated tropical & extreme tropical conditions.
Ecocentric System :
Chamber runs on the basis of Dewpoint control. Less deviations during short-term door opening & power failure. Less utility area Increased storage area. Low operating cost. Minimum power & water consumption. Reduced deviations. Less breakdowns due to standby refrigeration & humidification system. Due to PLC HMI & controlling through microprocessor based PID Temp. & RH operations the on time of both Temp. & RH system is reduced & efficiency of this system is increased with reduced energy consumption.







Salient Features : [Optional]
First time in India developed by SR Lab Instruments. Standby Refrigeration, Standby Humidification & Standby safety controller with auto changeover, In case of failure of the main system, To run the system for a longer with uninterrupted operation. Also USFDA approved 21 CFR compliance software, where in case of temp. & RH fluctuation, SMS & Email alert shall be sent to registered user nos and minute details of operation of the chamber can be downloaded weekly, Monthly or Quarterly.
Controller : [PLC with HMI TFT screen]
PID Action Provides Precise Temperature, Maintain Uniform
And Accurate Growth Environment, Timer, Alarm, Auto-Tuning
And Auto Start Stop Function. Technical Design and Construction
Silent Fan Motor to Maintain Uniform Temperature Over Temp.
and Current Protection Capability Ensure User Safety, Adjustable
Sliding Rack, Alert Unexpected Interruption Of Electrical Power
Or Unauthorized Change.
PLC-HMI Will rise gradually to reach the temp. in fixed time, remain in fixed temp. For Known period of time & then decreased
back to initial temp, in fixed time internal gradually. Stepwise
Programmable Controller with 50 or more programs Provides
Automatic Operation of Variable Temperature, Humidity And
Illumination Value, 100 Or More Cycles With Minimum 100
Sections in Each Cycle. Data logger with RS 485/232 port for
logging & transfer of data. Auto restart, power failure event
logging, memory protection up to 5-7 years. S.M.A.R.T/ SELF
diagnostics system. Technical Design and Construction Silent
Fan Motor to Maintain Uniform Temperature and Humidity Light
Bank System. Control system shall have audio/audiovisual
warning/alarm when set temperature/RH/ lighting is below or
above the set values. Ethernet communication port. Ambient
temp. monitoring. Protects Plants from Stress By Preventing
Sudden Change Of Light Strength Safety And Convenience.
Over Temperature and Current Protection Capability Ensure
User Safety, Adjustable Sliding Rack, Water Level Sensor Alert
Control System. Total no of output: 23 channels.
Alert Unexpected Interruption of Electrical Power or
Unauthorized Change.4 digit password lock
Refrigeration system : (Air Cooled)
Emerson /Danfoss/Tecumseh : make compressors, 2 stage cooling system for TEMP. & Humidity Control. All accessories such as HPCO/LPCO, Oil Separators, Dryers,Relays, OLP etc. of Danfoss/Tecumseh Emerson Make. CFC free refrigerant to be used to comply with the New Environmental Regulation. Will have suitable tonnage compressor for maintain temperature with lights on. Will have hermetically sealed condensing unit compressor coupled with Evaporation coil and condenser, safe guarded by time delay circuit. Cabinet shall be supplied with air-cooled condensing unit with extended compressor life and close temperature control. Cooling system will be split type to place condenser unit in a ventilated area to avoid hot air in growth chamber. Cooled air shall be delivered in such a way to reduce or eliminate condensation within media dishes and to provide maximum uniformity In different research vessels. This may be through silent and durable fan motors or through any kind of advanced air diffusers. Air Cooled Condenser with solenoid & LP to pass hot gas to the chamber for uniformity & long life of compressor for continuous usage. Phenolic coating for cooling & condensing coils (All weather proof).
Cabinet Construction & Insulation :
MOC of the Chamber will be: Double walled Chamber with Inner & outer
of GI duly pre painted PUF Insulation in between two walls.
Insulation of Walls, ceiling and floor :100 mm thick puff panels with puff in-place density of minimum of 1kg per cubic foot. Exterior metal surface of insulation puff panels or its equivalent shall be of minimum 0.5mm gauge GIPP. Interior metal surface of insulation puff panels or its equivalent shall be of minimum 0.5mm gauge GIPP. Exterior and interior metal surface shall not have metal-to- metal bonding.
Doors : Flush type doors of metal surfaces (as per above specification of metal) with proper puff insulation of above mentioned standard. Doors shall have magnetic snap-in perimeter gasket or its equivalent, self-closing cam lift gravity hinges or its equivalent, a posi-seal door closure or its equivalent. Door shall have key lockable latch handle. Doorjambs shall be made fiberglass reinforced plastic or any other suitable material compatible with constructed door.
Air Circulation :
The new walk in Chamber is provided with best air circulation to give better temperature uniformity this is achieved by using one set of motor & blower in baffle wall Compartment to push treated air in working chamber. This system gives better than 1 deg c. temperature uniformity in working chamber, evaporator, heaters & steam injection system is from back side, behind baffle wall, the air circulation blower is also fitted in this compartment only. This arrangement helps to mix steam and air together and pass well mixed air inside the working chamber. Due to above system the temperature and humidity (R.H.) in complete chamber almost remains equal.
Notch Cutting Device :
Digital Notch depth measuring device (0 to 12 mm) Suitable for notching plastics specimens for Izod & Charpy impact testes as per ASTM & ISO specification. The machine shall have gradual feed control mechanism with micrometer& standard notching tool. Motor driven, attached with constant profile tungsten carbide knife,”V” notch (45), Type A, Type B and Type C Comply to ASTM D 256. Knife Speed Regulator. Digital tool with zeroing having accuracy of 0.01mm division for measurement of notch depth
Temp. Range :
-40°C to +100°C , Resolution : 0.1°C
Temp. Fluctuation : ± 0.1°C to ± 0.5°C
Heating Rate : Normal.
Cooling Rate : Normal.
Safety Features :
2 minute compressor “on” delay timer to safeguard the compressor.
Compressors overload relay protector. Electronic low water level cut
off device to cut off the supply to boiler heater in case of low water
level. Safety Temp. Controller. MCB for mains.
Trays & Racks :
Removable SS perforated trays are provided, adjustable and will slide out for easy clean up. Assembled Stainless Steel racks to accommodate trays
Power :
Single phase 230V AC, 50Hz / Three phase 440V AC, 50Hz
Certification :
MSME AWARD FOR BEST MANUFACTURING
PRACTICES - 2023.
ISO/IEC 17025: 2017 / Electrical Safety IEC 61010.
ISO 45001:2018 / ISO 14001:2015 & WHO-GMP
ISO 9001:2015 certified Co. & CE marked product.
RoHS 2002/95/EC / UL (QAMS Certification).
US FDA Certification.
NSIC & MSME registered.
D&B registered (Financial Certification).
Documentation :
To comply with the documentation requirements, we provide
IQ, OQ & PQ protocols to be executed before taking the chamber
in to for regular use, and support the supply with the following
documents with the detail operational and service manual.
Standard Operating Procedure (SOP)
Operational Manual for Controller
IQ,PQ,OQ protocols certificate.
Calibration certificate of all controlling modules with traceable to
NABL / ERTL accredited labs.
Certificate of MOC.
Test Report of chamber prior to supply with mapping certificate.
wiring diagram for ease of service maintenance.