srlabinstruments.com

Burn in Chamber

Burn in Chamber
Model : SRL / BC-33

Application :

These equipments are used for climatic & durability tests of electrical & electronic components, corrosion test on mechanical assemblies, materials for simulated tropical & extreme tropical conditions.

Heating System :

Nichrome wire coiled or strip heaters or tubular heaters are used depending upon the design of the Burn-In-Chamber which may vary with the specific application and this should be conveyed to us by the user to ensure optimum performance parameters. A port hole of suitable dimension will be provided for giving power outside.

Construction :

MOC of the Chamber will be: Double walled Chamber with Inner & outer of GI duly pre painted, PUF Insulation in between two walls.

  • Flooring: 2 mm aluminum chequer plates.
  • The unit will be made of double walled leak proof metal door with lock, with a provision of opening door from inside
  • Forced air circulation for uniform temperature.
  • Insulation of Walls, ceiling and floor : : 60-80mm thick puff panels with puff in-place density of minimum of 1kg per cubic foot or its equivalent. Insulation with expended polystyrene or any equivalent material having R factor of minimum 34. Insulation material shall maintain its dimensional o 0 stability in an operating temperature range of 5 C to +60 C Insulation puff.
    Door : : Flush type doors of metal surfaces (as per above specification of metal) with proper puff insulation of above mentioned standard or its equivalent. Doors shall have magnetic snap-in perimeter gasket or its equivalent, self-closing cam lift gravity hinges or its equivalent, a posi-seal door closure or its equivalent. Door shall have key lockable latch handle with an inside safety release. Doorjambs shall be made fiberglass reinforced plastic or any other suitable material compatible with constructed door.

    Power Supply :

    Single Phase 230V AC, 50Hz / Three Phase 440V AC, 50Hz

    Air Circulation :

    The chamber is provided with best air circulation to give better temperature uniformity this is achieved by using one set of motor & blower in baffle wall Compartment to push treated air in w orking chamber. This system gives better than 1 deg c. temp. uniformity in working chamber. Uniform forced air circulate across the shelf via air diffusers on the top wall. Uniform horizontal or vertical airflow shall be provided for maintaining uniform temperature and RH in the room. System for adjustable forced air exchange shall be provided with up to 20 air exchanges per hour of fresh air to the room or its equivalent and an option to shut it down when not required. No direct radiation on the specimen

    Safety Features :

    2 minute compressor “on” delay timer to safeguard the compressor. Compressors overload relay protector. Electronic low water level cut off device to cut off the supply to boiler heater in case of low water level. Safety Temp. Controller. MCB for mains.

  • Unit will be provided with safety devices for temperature and humidity overshoot in case of malfunction.
  • Built-in temp. Deviation, audio/visual alarms. Safety thermostat for over shooting of temp. Safety circuit to cut off the whole system
  • OLP (Overload Protector) & Time delay Circuit for safety of compressors.
  • HRC fuses for compressors, Heaters & Mains.
  • Temp. Range :

    Temp Range : Above ambient 5 C to 100 C
    Accuracy : 0.2 C
    Uniformity : 0.2 C

    Certification :

    MSME Award For Best Manufacturing Practices 2023.
    ISO 9001:2015 Certified Co. D&B registered & CE mark product.
    ISO 14001:2015 & WHO-GMP
    ISO 45001:2018 & ISO/IEC 17025: 2017
    NSIC & MSME registered / ROHS Compliance / UL Compliance

    Documentation :

    To comply with the documentation requirements, we provide IQ, OQ & PQ protocols to be executed before taking the chamber in to for regular use, and support the supply with the following documents with the detail operational and service manual.
    Standard Operating Procedure (SOP)
    Operational Manual for Controller
    IQ,PQ,OQ protocols certificate.
    Calibration certificate of all controlling modules with traceable to NABL / ERTL accredited labs.
    Certificate of MOC.
    Test Report of chamber prior to supply with mapping certificate. wiring diagram for ease of service maintenance.